2D-INADEQUATE setup note

  1. Use inadgpqfsp sequence. (If use inadqf sequence, strong t1 noise and crossing artifacts will be shown.)
  2. Normally, use: d1 = 5s, ns >= 128, td1 ~64, and choose proper sw and o1.
  3. The sample concentration should be >= 400mg/ml, and an overnight expt should be enough.
  4. (Use INAD parameter folder, set in two gradient pulses, edasp, getprosol, rga, zg.)

Hongwei edited on 8/15/2023